Windowed DIA#

Summary#

Name

Pub date

Scan per cycle

Window size (m/z)

MS2 window features

MS1

MS2

Acquired

Demul.

Consecutive

Overlap (m/z)

DIA [1]

2004 Sep.

0

100

10

/

T

none

PAcIFIC [2]

2009 July

0

10

2.5

/

T

margin (1)

XDIA [4]

2010 Jan.

1

60

20

/

T

margin (1)

FT-ARM [5]

2012 Jan.

0

5

100

/

T

none

SWATH [6]

2012 Jan.

1

32

25

/

T

margin (1)

MSX [7]

2013 June

0/2

20

20

4

random

none

WiSIM-DIA [8]

2014

3

51

12

/

T

none

pSMART [10]

2014 Sep.

5

110

5,10,20

/

T

margin (1)

HRM [11]

2015 Feb.

1

19

24-222

/

T

margin (2)

SONAR [12]

2017 Sep.

200*

200*

24

2.5

T

scanning Q1

μDIA [14]

2018 July

1

120

9

3

T

one-third (3)

Staggered DIA [15]

2018 Dec.

2

20

20

10

T

one-half (10)

RTwinDIA [17]

2019 May

1

40

5

/

T

none

DIA-FAIMS [18]

2020 Feb.

1

49

13.7

/

T

margin (1)

PulseDIA [19]

2020 Sep.

1

24

5-29/11-71

/

reg. inter.

margin (1)

PASS-DIA [20]

2020 Oct.

5

75

2

/

T

none

HRMS1-DIA [21]

2020 Oct.

3

54

15

/

T

none

BoxCarmax [22]

2021 Feb.

1

30

10

2.5

reg. inter.

margin (1)

Scanning SWATH [23]

2021 Mar.

0

250*

10

2

T

scanning Q1

BoxCar DIA [24]

2021 July

4

24

20-583

/

reg. inter.

margin (1)

Zeno SWATH [25]

2022 Nov.

1

85

6-10

/

T

margin (1)

Narrow-window DIA [26]

2023

1

300

2

/

T

none

Note

  • Pub date
    • the date the method was publicly released by publisher

    • WiSIM-DIA is released as a Thermo Application Note

    • Narrow-window DIA is currently a pre-print

  • Scan per cycle
    • Number of finally recorded scans for MS1 or MS2 in one cycle

    • SONAR and Scanning SWATH are annotated by the numbers of bins at Q1 level

  • Window size (m/z)
    • Acquired is the real window size defined in acquisition settings for instrument, and the overlapped size is eliminate (e.g. if 21 m/z widnow size is used to ensure a 1 m/z overlap within adjacent windows, its effective size is 20 m/z)

    • Demul. is the computationally demultiplexed window size (so-called demultiplexing region in [15])

  • MS2 window features
    • Consecutive indicates if the MS2 windows in a cycle are defined in a continous manner (reg. inter. means regular interval)

    • Fixed indicates if all MS2 windows have the same size

    • Overlap indicates how the adjacent MS2 windows overlapped (though the adjacent windows might be not consecutive)

  • All window sizes included in this document have margin overlap subtracted

Individuals#


DIA#

back to table

Reference:

Full name

Data-Independent Acquisition

Authors

John D Venable et al.

Lab & Company

Yates Lab

Publication date

2004, Sep.

MS instrument

LTQ

MS1 spectra per cycle

0

MS2 spectra per cycle

100

MS2 window size

10 m/z

Demultiplexing region

/

Covered mass range

400-1400 m/z

Injection times

1

Cycle time

~35 s

MS2 window features

Consecutive & fixed window size & non-overlapping windows

  • Vanilla DIA scheme, with no MS1 scans, and a fixed 10 m/z window runs on pre-defined m/z range in a cycle without overlap.

PAcIFIC#

back to table

Reference:

Full name

Precursor Acquisition Independent From Ion Count

Authors

Alexandre Panchaud et al.

Lab & Company

Goodlett Lab

Publication date

2009 July

MS instrument

LTQ Orbitrap XL

MS1 spectra per cycle

0

MS2 spectra per cycle

100

MS2 window size

10 m/z

Demultiplexing region

/

Covered mass range

400-1400 m/z

Injection times

1

Cycle time

~35 s

MS2 window features

Consecutive & fixed window size & margin-overlapping windows (1 m/z)

  • 2.5 m/z window, and a 1 m/z overlap between adjacent windows.

  • The first MS2 scan in the first injection starts from a center mass of 401 m/z, and the last MS2 scan in last injection has a center mass of 1403.5 m/z.

  • Requires ~67 injections and ~5 days to finish the acquisition for one sample in 2009 on a LTQ Orbitrap XL instrument, and the effiency is improved to ~2 days with a faster LTQ Orbitrap Velos in 2011.

XDIA#

back to table

Reference:

Full name

extended DIA

Authors

Paulo C. Carvalho et al.

Lab & Company

Yates Lab

Publication date

2010, Jan.

MS instrument

LTQ-Orbitrap XL (+ETD)

MS1 spectra per cycle

1

MS2 spectra per cycle

60

MS2 window size

20 m/z

Demultiplexing region

/

Covered mass range

400-1000 m/z

Injection times

1

Cycle time

/

MS2 window features

Consecutive & fixed window size & margin-overlapping windows (1 m/z) & two scans for each window (one for ETD, one for ETD+CID)

  • Each acquisition window is scaned twice, by using ETD in the first scan and followed by ETD+CAD in the second scan for ion dissociation.

FT-ARM#

back to table

Reference:

Full name

Fourier Transform-All Reaction Monitoring

Authors

Chad R. Weisbrod et al.

Lab & Company

Bruce Lab

Publication date

2012, Jan.

MS instrument

LTQ-FT and LTQ Orbitrap

MS1 spectra per cycle

0

MS2 spectra per cycle

5

MS2 window size

100 m/z

Demultiplexing region

/

Covered mass range

500-1000 m/z

Injection times

1

Cycle time

~5.45 s

MS2 window features

Consecutive & fixed window size & non-overlapping windows

SWATH#

back to table

Reference:

Full name

Sequential Window Acquisition of all THeoretical (mass spectra)

Authors

Ludovic C. Gillet et al.

Lab & Company

Aebersold Lab

Publication date

2012, Jan.

MS instrument

TripleTOF 5600

MS1 spectra per cycle

1

MS2 spectra per cycle

32

MS2 window size

25 m/z

Demultiplexing region

/

Covered mass range

400-1200 m/z

Injection times

1

Cycle time

~3.3 s

MS2 window features

Consecutive & fixed window size & margin-overlapping windows (1 m/z)

MSX#

Reference:

Full name

MSX

Authors

Jarrett D Egertson et al.

Lab & Company

MacCoss Lab

Publication date

2013, June

MS instrument

Q Exactive

MS1 spectra per cycle

0 (or 2 for MS1 quantification)

MS2 spectra per cycle

20 (5 discrete sub-windows in each scan)

MS2 window size

20 m/z (5 * 4 m/z sub-window for ion accumulation)

Demultiplexing region

4 m/z

Covered mass range

500-900 m/z

Injection times

1

Cycle time

~3.5 s

MS2 window features

5 sub-windows in each scan are randomly picked & fixed window size & non-overlapping windows

  • Actual center position of Q1 in the first 4 m/z segment is 502.4783 m/z with a window width of 4.002 m/z, agreeing with m/z distribution of precursor residue combinations.

WiSIM-DIA#

back to table

Reference:

Full name

Wide Selected-Ion Monitoring DIA

Authors

Reiko Kiyonami et al.

Lab & Company

Thermo Fisher Scientific GmbH

Publication date

2014

MS instrument

Orbitrap Fusion Tribrid

MS1 spectra per cycle

3

MS2 spectra per cycle

51

MS2 window size

12 m/z

Demultiplexing region

/

Covered mass range

400-1004 m/z

Injection times

1

Cycle time

~3.6 s

MS2 window features

Consecutive & fixed window size & non-overlapping windows

  • MS2 scans are performed in ion trap in parallel with MS1 scan in Orbitrap

pSMART#

back to table

Reference:

Full name

pSMART

Authors

Amol Prakash et al.

Lab & Company

Thermo Fisher Scientific GmbH

Publication date

2014, Sep.

MS instrument

Q Exactive

MS1 spectra per cycle

5

MS2 spectra per cycle

110

MS2 window size

5 m/z window for 400-800 m/z, 10 m/z window for 800-1000 m/z, 20 m/z window for 1000-1200 m/z

Demultiplexing region

/

Covered mass range

400-1200 m/z

Injection times

1

Cycle time

~26 s

MS2 window features

Consecutive & variable window sizes & margin-overlapping windows (1 m/z)

  • In each cycle, 5 HR/AM MS1 spectra are acquired every ~5 s.

HRM#

back to table

Reference:

Full name

Hyper Reaction Monitoring

Authors

Roland Bruderer et al.

Lab & Company

Biognosys AG

Publication date

2015, Feb.

MS instrument

Q Exactive

MS1 spectra per cycle

1

MS2 spectra per cycle

19

MS2 window size

Variable (24-222 m/z)

Demultiplexing region

/

Covered mass range

400-1220 m/z

Injection times

1

Cycle time

~3.5 s

MS2 window features

Consecutive & variable window sizes & margin-overlapping windows (2 m/z)

SONAR#

back to table

Reference:

Full name

SONAR

Authors

  1. Arthur Moseley et al.

Lab & Company

Waters Corporation

Publication date

2017, Sep.

MS instrument

Xevo G2-XS

MS1 spectra per cycle

200

MS2 spectra per cycle

200

MS2 window size

24 m/z

Demultiplexing region

2.5 m/z

Covered mass range

400-900 m/z

Injection times

1

Cycle time

~1 s

MS2 window features

Consecutive & fixed window size & intra-overlapping windows (21.5 m/z at Q1 bin level)

μDIA#

back to table

Reference:

Full name

microDIA

Authors

Michael R. Heaven et al.

Lab & Company

Norris Lab

Publication date

2018, July

MS instrument

impact II

MS1 spectra per cycle

1

MS2 spectra per cycle

120

MS2 window size

9 m/z

Demultiplexing region

3 m/z

Covered mass range

400-1115 m/z

Injection times

1

Cycle time

~3.4 s

MS2 window features

Consecutive & fixed window size & intra-overlapping windows (3 m/z)

Staggered DIA#

back to table

Reference:

Full name

microDIA

Authors

Brian C. Searle et al.

Lab & Company

MacCoss Lab

Publication date

2018, Dec.

MS instrument

Q Exactive

MS1 spectra per cycle

2

MS2 spectra per cycle

20

MS2 window size

20 m/z

Demultiplexing region

10 m/z

Covered mass range

490-900 m/z

Injection times

1

Cycle time

~2.5 s

MS2 window features

Consecutive & fixed window size & inter-overlapping windows (10 m/z)

  • In two-step cycles, odd-numbered cycle covers 500-900 m/z and even-numbered cycle covers 490-890 m/z.

RTwinDIA#

back to table

Reference:

Full name

RT windowed DIA

Authors

Wenxue Li et al.

Lab & Company

Liu Lab

Publication date

2019, May

MS instrument

Orbitrap Fusion Lumos

MS1 spectra per cycle

1

MS2 spectra per cycle

40

MS2 window size

5 m/z

Demultiplexing region

/

Covered mass range

400-1000 m/z

Injection times

1

Cycle time

~3.2 s

MS2 window features

Consecutive & fixed window size & non-overlapping windows

  • MS1 covers 350-1650 m/z. MS2 in 3 blocks: 400-600 m/z for RT range 0-50%; 600-800 m/z for RT range 50-75%; 800-1000 m/z for RT range 75-100%.

DIA-FAIMS#

back to table

Reference:

Full name

DIA-FAIMS

Authors

Dorte B. Bekker-Jensen et al.

Lab & Company

Olsen Lab

Publication date

2020, Feb.

MS instrument

Orbitrap Exploris 480

MS1 spectra per cycle

1

MS2 spectra per cycle

49

MS2 window size

13.7 m/z (15k2s method)

Demultiplexing region

/

Covered mass range

361-1032.3 m/z (15k2s method)

Injection times

1

Cycle time

~2.3 s (15k2s method)

MS2 window features

Consecutive & fixed window size & margin-overlapping windows (1 m/z)

PulseDIA#

back to table

Reference:

Full name

PulseDIA

Authors

Xue Cai et al.

Lab & Company

Guo Lab

Publication date

2020, Sep.

MS instrument

Q Exactive HF

MS1 spectra per cycle

1

MS2 spectra per cycle

24

MS2 window size

Variable 5-29 m/z for 4 injections, and variable 11-71 m/z for 2 injections

Demultiplexing region

/

Covered mass range

400-1200 m/z

Injection times

2-5

Cycle time

~2 s

MS2 window features

Discrete & variable window size & margin-overlapping windows (1 m/z)

  • Each cycle has 1 MS1 scan and 24 MS2 scans. MS1 covers 390-1210 m/z, and MS2 from all injections covers 400-1200 m/z.

  • The MS2 covered 800 m/z are split into 24 segments, in which 20 segments with 20 m/z width cover 400-800 m/z, and the other 4 segments cover 800-860, 860-940, 940-1060, 1060-1200 m/z.

  • Each segment is equally divided into n parts, where n is the injection number, and 1 m/z overlap is added when determining the window size.

  • Four window schemes are defined: 2-injection has window sizes of 11, 31, 41, 61, and 71 m/z; 3-injection has window sizes of 8, 21, 28, 41, and 48 m/z; 4-injection has window sizes of 6, 16, 21, 31, and 36 m/z; 5-injection has window sizes of 5, 13, 17, 25, and 29 m/z.

PASS-DIA#

back to table

Reference:

Full name

Precursor ion And Small Slice-DIA

Authors

Dong-Gi Mun et al.

Lab & Company

Pandey Lab

Publication date

2020, Oct.

MS instrument

Q Exactive HF

MS1 spectra per cycle

5

MS2 spectra per cycle

75

MS2 window size

2 m/z

Demultiplexing region

/

Covered mass range

350-1400 m/z

Injection times

7

Cycle time

/

MS2 window features

Consecutive & fixed window size & non-overlapping windows

  • Each injection covers 150 m/z. Each cycle has 75 MS2 scans with a fixed 2 m/z window size, and MS1 scan is acquired per 15 MS2 scans.

HRMS1-DIA#

back to table

Reference:

Full name

High-Resolution MS1-based quantitative DIA

Authors

Yue Xuan et al.

Lab & Company

Thermo Fisher Scientific GmbH

Publication date

2020, Oct.

MS instrument

Q Exactive HF

MS1 spectra per cycle

3

MS2 spectra per cycle

54

MS2 window size

15 m/z

Demultiplexing region

/

Covered mass range

400-1210 m/z

Injection times

1

Cycle time

~5.23 s

MS2 window features

Consecutive & fixed window size & non-overlapping windows

  • Each cycle comprises 3 MS1 full scans and 54 sequential MS2 scans. MS1 full scans are inserted before the 1st, 18th, 35th MS2 scans. Each MS1 full scan covers whole mass range of 400-1210 m/z.

BoxCarmax#

back to table

Reference:

Full name

BoxCarmax

Authors

Barbora Salovska et al.

Lab & Company

Liu Lab

Publication date

2021, Feb.

MS instrument

Orbitrap Fusion Lumos

MS1 spectra per cycle

1

MS2 spectra per cycle

30

MS2 window size

10 m/z (4 * 2.5 m/z sub-windows)

Demultiplexing region

/

Covered mass range

357-1197 m/z

Injection times

4

Cycle time

~2.8 s

MS2 window features

Discrete & fixed window size & margin-overlapping windows (1 m/z)

Scanning SWATH#

back to table

Reference:

Full name

Scanning SWATH

Authors

Christoph B. Messner et al.

Lab & Company

Ralser Lab

Publication date

2021, Mar.

MS instrument

TripleTOF 6600

MS1 spectra per cycle

0

MS2 spectra per cycle

250

MS2 window size

10 m/z

Demultiplexing region

2 m/z

Covered mass range

400-900 m/z

Injection times

1

Cycle time

~0.52s (5 min gradient), ~0.31s (1 min gradient), ~0.28s (0.5 min gradient)

MS2 window features

Consecutive & fixed window size & intra-overlapping windows (8 m/z at Q1 bin level)

BoxCar DIA#

back to table

Reference:

Full name

BoxCar DIA

Authors

Pavel Sinitcyn et al.

Lab & Company

Cox Lab

Publication date

2021, July

MS instrument

Orbitrap Fusion Lumos

MS1 spectra per cycle

4

MS2 spectra per cycle

24

MS2 window size

Variable (20-583 m/z)

Demultiplexing region

/

Covered mass range

350-1650 m/z

Injection times

1

Cycle time

~3.3 s

MS2 window features

Consecutive & variable window size & margin-overlapping windows (1 m/z)

  • “BoxCar DIA” instrument settings were extracted from “20201006_LU1_DaIt_Aurora5_HEK1_BOXCAR_SWATH_2ul.raw” from PXD022589 accompanying with MaxDIA paper.

  • Each cycle has 1 MS1 full scan, 3 tSIM scans, and 24 MS2 scans sequentially arranged. MS1 full scan covers 350-1650 m/z. 3 tSIM scans are interspersed with each other like those in BoxCar DDA, start from 350-385, 384-412, and 411-434 m/z, respectively, and end at 898-966, 965-1068, 1067-1650 m/z, respectively.

Zeno SWATH#

back to table

Reference:

Full name

Zeno SWATH

Authors

Ziyue Wang et al.

Lab & Company

Ralser Lab

Publication date

2022 Nov.

MS instrument

ZenoTOF 7600

MS1 spectra per cycle

1

MS2 spectra per cycle

85

MS2 window size

6-10 m/z

Demultiplexing region

/

Covered mass range

400-903 m/z

Injection times

1

Cycle time

/

MS2 window features

Consecutive & variable window size & margin-overlapping windows (1 m/z)

Narrow-window DIA#

back to table

Reference:

Full name

Narrow-window DIA

Authors

Ulises H. Guzman et al.

Lab & Company

Olsen Lab

Publication date

2023

MS instrument

Orbitrap Astral

MS1 spectra per cycle

1

MS2 spectra per cycle

300

MS2 window size

2 m/z

Demultiplexing region

/

Covered mass range

380-980 m/z

Injection times

1

Cycle time

/

MS2 window features

Consecutive & fixed window size & non-overlapping windows