Full scan DIA#
Summary#
Name |
Pub date |
Instrument |
Lab & Co. |
|---|---|---|---|
Shotgun CID [1] |
2003 June |
TOF |
Goodlett Lab |
MSE [2] |
2005 Mar. |
Q-TOF Ultima |
Waters Corporation |
AIF [3] |
2010 July |
Orbitrap |
Mann Lab |
UDMSE [4] |
2013 Dec. |
Synapt G2-S |
Tenzer Lab |
Note
Pub date: the date the method was publicly released by publisher
Instrument: initially developped on which instrument
Lab & Co.: lab or company (the last corresponding author)
Individuals#
Shotgun CID#
back to table
- Reference:
Full name |
Shotgun CID |
Authors |
Samuel Purvine et al. |
Lab & Company |
Goodlett Lab |
Publication date |
2003, June |
MS instrument |
TOF-MS |
Covered mass range |
400-1200 m/z |
MSE#
back to table
- Reference:
Quantitative proteomic analysis by accurate mass retention time Pairs. 10.1021/ac048455k
Full name |
MSE |
Authors |
Jeffrey C. Silva et al. |
Lab & Company |
Waters Corporation |
Publication date |
2005, Mar. |
MS instrument |
Q-TOF Ultima |
Covered mass range |
300-2000 m/z |
AIF#
back to table
- Reference:
Proteomics on an Orbitrap benchtop mass spectrometer using all-ion fragmentation. 10.1074/mcp.M110.001537
Full name |
All-Ion Fragmentation |
Authors |
Tamar Geiger et al. |
Lab & Company |
Mann Lab |
Publication date |
2005, Mar. |
MS instrument |
Exactive (single Orbitrap) |
Covered mass range |
300-1600 m/z (MS1) and 100-1600 m/z (MS2) |
UDMSE#
back to table
- Reference:
Drift time-specific collision energies enable deep-coverage data-independent acquisition proteomics. 10.1038/nmeth.2767
Label-free quantification in ion mobility-enhanced data-independent acquisition proteomics. 10.1038/nprot.2016.042
Full name |
ultra-definition MSE |
Authors |
Ute Distler et al. |
Lab & Company |
Tenzer Lab |
Publication date |
2013, Dec. |
MS instrument |
Synapt G2-S |
Covered mass range |
350-2000 m/z |
- Increased fragmentation efficiency by optimizing CE values according to TWIMS ion mobility
CE 17 eV for ion mobility bins 0-20
CE ramp from 17 eV to 45 eV for ion mobility bins 20-110
CE ramp from 45 eV to 60 eV for ion mobility bins 110-200